Troubleshooting
Error
E023
Cleaning Solution
E024
Input active
E026
IC LM75
E028
Signal output open
E030
EEprom Frontend
E031
Calibration Recout
E032
Wrong Frontend
E039
Alarm high pH
E040
Alarm low pH
E049
Power-on
E050
Power-down
E065
DPD / Buffer
E067
Cleaning Solution
72
Description
Corrective action
– refill cleaning solution, see
– See If Fault Yes is programmed in Menu
5.3.4, p. 102
– call service
– check wiring on signal outputs 1 and 2
– call service
– call service
– call service
– check process
– check programmed value
– check process
– check programmed value
– none, normal status
– none, normal status
– Operating display, upper status line: The
number next to the triangle, indicates
the remaining reagents in%. Refill
reagents on time. See
Reagents, p. 46
– Operating display, lower status line: The
number next to the triangle, indicates
the remaining cleaning solution in%.
Refill cleaning solution on time.
6.3.,
46
5.3.1.6, p. 96
5.3.1.6, p. 96
Refill or replace