AMI Codes-II CC
Maintenance
Zero
Process pH
52
A zero is automatically done before each measurement.
Use a Chematest 25 photometer (or equivalent) to determine the
sample pH value. Insert the electrode through a hole in the con-
stant head cover into the constant head.
NOTICE: Make sure your reference instrument is calibrated
correctly!
Maintenance
Calibration
Calibration
Service
Simulation
Set Time 01.01.05 16:30:00
Cleaning
Enter the correct value
with the [
] or [
key.
Process pH
Current Value
Offset
Slope
x.xx mV
Calibration successful
Possible error messages see
3.1
Enter
]
3.1.2.5
7.70 pH
Enter
y mV
Calibration Errors, p.
Calibration
Free av. chlorine
Total chlorine 2
Process pH
Standard pH
Process pH
Current Value
7.78 pH
Offset
Process Value
7.78 pH
Save
<Enter>
Process pH
Current Value
7.78 pH
Offset
Process Value
7.70 pH
Save
<Enter>
Process pH
Current Value
7.70 pH
Offset
Process Value
7.70 pH
Save
<Enter>
68.
A-96.250.581 / 220616
3.1.2
3.1.2.4
x mV
3.1.2.4
x mV
3.1.2.5
y mV