AMI Codes-II CC
Maintenance
Prepare
Oxycon
On-line Buffer
Prepare
Oxycon
On-line KI
A-96.250.581 / 220616
4 Close the canister with the screw cover and tighten it well.
5 Mix the demineralized water and the reagents well.
6 Put the canister [E] into the holder [H].
7 Remove the screw cover and insert the suction lance [C] com-
bined with level detector [B] and tighten the screw cover [A].
1 Rinse the canister [F] labelled "OXYCON ON LINE Buffer" solu-
tion with demineralized water.
2 Fill the canister up to the 2 liter mark [D] with demineralized wa-
ter.
3 Slowly pour the content of one bag of buffer Oxycon On-line
Buffer into the canister.
Avoid splashing!
4 Close the canister with the screw cover and tighten it well.
5 Mix the demineralized water and the reagents well.
6 Put canister [F] into holder [H].
7 Remove the screw cover and insert the suction lance [C] com-
bined with level detector [B] and tighten the screw cover [A].
1 Rinse the canister [G] labelled "OXYCON ON LINE KI" solution
with demineralized water.
2 Fill the canister up to the 2 liter mark [D] with demineralized wa-
ter.
3 Slowly pour the content of one bag of Oxycon On-line KI into
the canister.
Avoid splashing!
4 Close the canister with the screw cover and tighten it well.
5 Mix the demineralized water and the reagents well.
6 Put canister [G] into holder [H].
7 Remove the screw cover and insert the suction lance [C] com-
bined with level detector [B] and tighten the screw cover [A].
8 Fill reagent system. See
Fill or Flush Reagent System, p. 38