AMI Codes-II CC
Product Description
G
G
J
fac
tc 1
1
2
A
Reaction time for measuring tc 2
B
Flush time
C
Overall measurement time
D
Flush time end
E
fac measurement
G
Valve 1 for adding OXYCON ON-LINE KI
J
Valve 2 for closing photometer inlet
fac
Free available chlorine
fac
fac fac
tc2
0
10
F
A-96.250.581 / 220616
tc 2
4
3
A
C
The programmed measuring intervals of tc2 [F] and fac [G] may
overlap each other.
The tc2 measurement has priority and therefore always takes place
in the programed measuring interval [F]. The fac measurement is
filled in the gaps [H] of the overall measurement and may be shifted
(orange arrow) in respect to programmed interval. If 2 fac measure-
ments are programmed during a overall measurement one of them
is cancelled (red arrow).
20
H
J
5
B
tc1
Total chlorine 1
tc2
Total chlorine 2
OXYCON ON-LINE DPD
OXYCON ON-LINE Buffer
OXYCON ON-LINE KI
Zero point measurement
Sample measurement
Peristaltic pump active for a short
time
30
40
G
fac
D
6
7
E
50
min
8
min
60
15