OPERATION

Basic Operation

Basic Operation
4-2
RC-6 PLUS
WARNING
1. Failure to load and install the rotor in accordance with the instructions in the
rotor operating guide could result in damage to the centrifuge. The rotor cover
must be on and locked in place and the rotor must be locked on to the drive
spindle.
2. Use specified Thermo Fisher Scientific rotors ONLY.
3. Do not incline or move the instrument while the rotor is spinning.
4. Do not place any object on the instrument or lean on the instrument while the
rotor is spinning.
This section describes how to perform a basic run, how to set run conditions on
the RUN SCREEN and how to use the cursor keys.
1. Set the main power switch ON.
2. Open the chamber door.
3. Install the rotor:
Wipe the rotor centerhole and spindle surfaces clean to reduce the chance
of the rotor sticking to the spindle.
If applicable, make sure the rotor cover is installed.
Place the rotor on the drive spindle and lock it in place turning the rotor
locking knob counterclockwise.
4. Close the chamber door.
5. Set the Run Screen as follows.