INSTALLATION
Electrical Requirements
Electrical
Requirements
2-4
RC-6 PLUS
Figure 2-2.
Leveling the Centrifuge
CAUTION
The centrifuge can be damaged if it is connected to the wrong voltage, or if it is
connected to a line voltage that varies more than plus/minus 10% of its nominal
value. Check the voltage before plugging the centrifuge into any power source.
Thremo Fisher Scientific is not responsible for improper installation.
The appropriate power source must be available to plug in the centrifuge. The
power supply and outlet type required varies, depending on the rating on the
nameplate. Check the nameplate on the back panel of the centrifuge to
determine the electrical configuration of your centrifuge, which should be one of
the following:
Table 2-2.
Electrical Requirements
NAME PLATE RATING
AC208/220/230/240
50/60Hz 21-24A single phase
AC200 50/60Hz
25A single phase
AC400 50/60Hz
2x15A 3 phase
SUPPLY REQUIRED
AC208/220/230/240 50/60Hz 30A single
phase (25A single phase may be
acceptable depending on you local
electrical codes).
AC200 50/60Hz 30A single phase
AC400 50/60Hz 2x25A 3 phase