Thermo Scientific RC 6 Plus Operation Instructions Manual page 47

Superspeed
Hide thumbs Also See for RC 6 Plus:
Table of Contents

Advertisement

Thermo Fisher Scientific
Step
Key operation
7
Check the RTC
setting on the
RUN SCREEN
and press the
START key.
1. The RUN mode indicator on the panel turns as follows in RTC operation.
Figure 4-8. Operating Mode (RTC)
2. Note that the RTC setting is not available in the following cases:
i.
The time setting on the RUN SCREEN is "HOLD" (continuous run).
Change the run time setting from "HOLD" to a desired numeric
value.
ii.
The start time has already passed.
Change the setting so that the start time will be later than the
current time.
iii. The start time is 20 days or more from the current time.
Change the setting so that the start time will be within 20 days.
RTC (Real Time Control) Operation
Screen display and notices
Perform
operation
Procedure.
Note that the run time setting cannot be changed
after setting the RTC operation. Clear the RTC
setting and then reset the run time if necessary.
The centrifuge turns to "DELAY" mode by pressing
the START key and wait until the set time. The
centrifuge automatically starts operation at the set
time and keeps operation during the designated
time.
The RUN mode indicator on the message display
turns to "DELAY".
OPERATION
according
to
Operating
RC-6 PLUS
4-19