2. Turn off the POWER switch of the centrifuge and the distribution board of
your centrifuge room.
3. Remove the two screws from the lower portion of the front cover. Remove
the front cover by pulling the lower portion of the front cover forward and
downward. The upper portion of the front cover is not secured with screws.
4. Raise the weights of the two door locks at the front and secure them with
adhesive tape.
Figure 4-9.
Opening the Door
5. Release the door handle and open the door slowly.
Check that the rotor has stoped completely. If the rotor is rotating, close the
door immediately.
Emergency Recovery from Power Failure
RC-6 PLUS
OPERATION
4-25