OPERATION
Displaying and Setting RCF
Displaying and
Setting RCF
4-20
RC-6 PLUS
3. Clear the RTC setting and then reset the run time if it is necessary to
change the run time setting after setting the RTC operation.
4. Recall the saved run conditions and enter the RTC setting when performing
the programmed operation (including the step-mode operation) and the
RTC operation in combination. The centrifuge automatically computes the
total run time of all steps in the programmed operation and also the start
time for RTC operation. Note that the saved run conditions cannot be
recalled after setting the RTC operation.
5. Press the STOP key to stop the operation. The RTC operation is stopped
and the rotor stops.
6. When performing the RCF operation and the RTC operation in combination,
enter the RCF setting first then enter the RTC setting. (See Displaying and
Setting RCF.)
7. When performing the ω
enter the ω
2
T setting first then enter the RTC setting. (See Displaying and
ω
2
Setting
T.)
The RC-6 Plus Superspeed centrifuge, has in its internal memory, data
representing the maximum radii of all available rotors. Based on this data, the
centrifuge automatically computes relative centrifugal force (RCF) values from
set speed, or speed from set RCF values, and then displays the result on the
control panel. This section explains how to use this RCF displaying and setting
capability of the centrifuge.
2
T operation and the RTC operation in combination,
Thermo Fisher Scientific