Thermo Scientific RC 6 Plus Operation Instructions Manual page 70

Superspeed
Hide thumbs Also See for RC 6 Plus:
Table of Contents

Advertisement

TROUBLESHOOTING
6-4
RC-6 PLUS
Table 6-3.
User Corrected Problems
Symptom
Run conditions cannot be
set or recalled.
Recalled run conditions
are changed.
Rotor is not cooled.
Cause Corrective
The rotor is still rotating.
The battery to back up the
program memory is dead.
The ambient temperature is
too high.
A heat-producing device
such as a refrigerator or a
generator
is
near
the
centrifuge.
The condenser fins are
clogged with dust.
Action
Set
or
recall
the
run
conditions when the rotor
stops completely.
Set the run conditions again
and keep the centrifuge
turned on for 10 hours to
recharge the battery.
Lower
the
room
temperature using an air
conditioner or lower the
speed
non-air-conditioned
environments.
Relocate
the
heat-producing device to
another place or contact
or
Thermo Fisher Scientific
your local representative
of
Thermo Fisher Scientific
products
to
ask
for
relocation
of
the
centrifuge.
Clean the condenser fins
according
to
the
procedure
specified
on
page 5-3.
Thermo Fisher Scientific
in