CAUTION
Do not pour any solution such as water, detergent or disinfectant directly into
the rotor chamber. Otherwise, the bearings of the drive unit may corrode or
deteriorate.
Before relocating the centrifuge, remove the rotor from the rotor chamber to
avoid damage to the drive shaft.
Always keep the liquid crystal panel in a visible position while the POWER
switch is turned on, so that you can check the current operating state (running
or stopping).
Safety During Installation and/or Maintenance
.
DANGER
When servicing the centrifuge, be sure to turn off the POWER switch, turn off
the distribution board of your centrifuge room, and then wait for at least three
minutes before removing covers or tables from the centrifuge to avoid electrical
shock hazards.
WARNING
Level the centrifuge by using the four level adjusters and secure them
completely. Improper securing can cause significant movement of the
centrifuge in the event of a rotor disengagement.
Installation or relocation of your centrifuge must be done by Thermo Fisher
Scientific or a Thermo Fisher Scientific representative. Contact Thermo Fisher
Scientific or your local representative of Thermo Fisher Scientific products
CAUTION
Avoid places that are exposed to ultraviolet rays for operation or storage of the
centrifuge. Otherwise, the covers can be discolored and the coating can be
peeled off easily. If installation in such a place is unavoidable, cover the
centrifuge with a cloth after operation to protect from ultraviolet rays.
Electrical Safety
WARNING
Your centrifuge must be grounded properly to avoid electrical shock hazards.
Thermo Fisher Scientific
RC-6 PLUS
iii