INSTALLATION
Chapter 2
Installation or relocation of your centrifuge must be done by an authorized
Thermo Fisher Scientific representative. Contact Thermo Fisher Scientific or
your local representative of Thermo Fisher Scientific products.
After you receive your centrifuge, inspect it for damage before using it. The
RC-6 PLUS centrifuge must be installed in a location that meets all of the
electrical, location, and environment requirements that are specified below and
on the following pages. Installation instructions begin on page 2-3.

Inspection

As soon as you receive your RC-6 PLUS Superspeed Centrifuge carefully
inspect it for any shipping damage that may have occurred. If you find any
damage, please report it immediately to the transportation company and file a
damage claim, then notify Kendro. If any parts are missing, contact one of the
Kendro district offices or the local representative of Thermo Fisher Scientific
products. You will find a list of offices on the back cover of this manual.
DANGER
When servicing the centrifuge, be sure to turn off the POWER switch, turn off
the distribution board of your centrifuge room, and wait for at least three
minutes before removing covers or tables from the centrifuge to avoid electrical
shock hazards.
Identify the
Installation Site
WARNING
For operator safety, maintain a 30-cm "clearance envelope" around the
instrument and keep out of that area while the rotor is spinning. Do not store
dangerous substances capable of developing flammable or explosive vapors in
the clearance envelope.
Thermo Fisher Scientific
RC-6 PLUS
2-1