OPERATION
RTC (Real Time Control) Operation
4-18
RC-6 PLUS
Step
Key operation
5
Enter the desired
date and time
(month, day, hour
and minutes) using
the cursor keys and
the ten-key numeri-
cal pad. Press the
ENTER key.
6
Select Yes or No in
response to the
prompt.
Screen display and notices
The range for "hour" setting is from 0 to 23
(24-hour display).
Do not enter any date and time that passed
the current time.
Set a proper stop time considering the
centrifugation time so that the start time will be
later than the current time.
It is not possible to set an operation that will
start 20 days or more from the current time.
When selecting "Yes":
The screen turns to the RUN SCREEN and the
RTC time is displayed.
RTC appears on the message indicator.
When selecting "No", the display turns to the
. Enter the desired setting
screen of Step 2
again.