OPERATION
RTC (Real Time Control) Operation
4-18
RC-6 PLUS
Step
Key operation
5
Enter the desired
date and time
(month, day, hour
and minutes) using
the cursor keys and
the ten-key numeri-
cal pad. Press the
ENTER key.
6
Select Yes or No in
response to the
prompt.
Screen display and notices
•
The range for "hour" setting is from 0 to 23
(24-hour display).
•
Do not enter any date and time that passed
the current time.
Set a proper stop time considering the
centrifugation time so that the start time will be
later than the current time.
•
It is not possible to set an operation that will
start 20 days or more from the current time.
•
When selecting "Yes":
•
The screen turns to the RUN SCREEN and the
RTC time is displayed.
•
RTC appears on the message indicator.
•
When selecting "No", the display turns to the
. Enter the desired setting
screen of Step 2
again.
Thermo Fisher Scientific