Setting RUN Conditions
Thermo Fisher Scientific
Table 4-1.
Step
Key Operation
1
Press the ROTOR key on the
KEY BOARD.
Enter the rotor number to be
used with the ten-key numerical
pad and press the NTER key.
(The screen turns to the next
page by pressing the cursor key.)
The screen returns to the RUN
SCREEN.
1. When setting two or more run conditions, there is no need to press the
ENTER key after each setting. The set value is entered by pressing a cursor
key and the cursor moves to the new item in ready-to-enter state.
2. If changing the TIME setting during continuous operation (HOLD), enter a
value adding the desired remaining time to the elapsed time. For example,
to stop the operation after 1 hour and 30 minutes when 5 hours have
elapsed in continuous operation, turn the TIME display to ready-to-enter
state by pressing a cursor key and enter as follows.
The table below exemplifies how to set run conditions such as rotor speed, run
time and rotor temperature.
Basic Operation
Screen Display and Notices
RC-6 PLUS
OPERATION
4-5