Centrifugeaccessories - Thermo Scientific RC 6 Plus Operation Instructions Manual

Superspeed
Hide thumbs Also See for RC 6 Plus:
Table of Contents

Advertisement

Centrifuge
Accessories
Thermo Fisher Scientific
***
After the centrifuge system has reached equilibrium. Ambient air temperature at the centrifuge
air inlets must be between 2°C to 40°C with maximum relative humidity of 80% at 31°C, linearly de-
creasing to 50% at 40°C. If the ambient air temperature is above 25°C, the centrifuge may not main-
tain low temperatures at high speeds, therefore, avoid areas near heat sources (for example, direct
sunlight, heating pipes and radiators). Also, avoid close grouping of centrifuges or other heat-pro-
ducing laboratory equipment.
****
As measured with SE-12 rotor at 21,000 rpm (with compressor turned on) 3 feet from the front
of the unit and four feet from the floor.
*****
For the SS-34 rotor spinning 20,000 rpm at 4°C, after it has reached equilibrium. Other rotors,
speeds, and temperatures cause the heat output to vary.
The following items are provided with the centrifuge:
Table 1-2.
Catalog No.
Description
50105750
US Instruction manual
50105751
EU Instruction manual
Level
Rear duct assembly (To be mounted to the rear cover at
installation)
65937
Silicon grease
61556
Lubricant grease
Rotor cleaning bar
INTRODUCTION & DESCRIPTION
Centrifuge Accessories
RC-6 PLUS
1-5