Table 4-3.
Temperature (TEMP)
4 degrees centigrade
The display turns to
ready-to-enter state.
The cursor is blinking at the single
digit position.
The temperature setting is "4
degrees centigrade".
-20 degrees centigrade to 40
degrees centigrade: in increments
of 1 degree centigrade
Thermo Fisher Scientific
Note
Keep the chamber door closed after the rotor has been removed to inhibit
the formation of condensation on the chamber walls.
To repeat the run with the same speed, time and temperature parameters,
install the rotor, close the chamber door, press START and then press ENTER.
Speed, Time and Temperature Parameters
Acceleration Rate (ACCEL)
9
The display turns to
ready-to-enter state.
The acceleration rate setting is
"9".
1 to 9
OPERATION
Basic Operation
Deceleration Rate (DECEL)
7
The display turns to
ready-to-enter state.
Enter "0" for selecting free coast.
The deceleration rate setting is
"7".
1 to 9 and 0 for free coast