12-Lead Print And Display Options - ZOLL Propaq MD Operator's Manual

Defibrillator
Hide thumbs Also See for Propaq MD:
Table of Contents

Advertisement

or a 3 x 4 Cabrera format:
The Propaq MD unit stores a minimum of 32 12-Lead snapshots in a separate log. Once 32 12-
lead snapshots are stored, the oldest snapshot in the log is overwritten by subsequent snapshots.
Note that if a patient data log is full, you will not be able to store12-lead snapshots. The 12-lead
snapshots are erased when you clear the log.
When you are done viewing and printing the 12-Lead waveform traces, press the Exit 12 quick
access key (

12-Lead Print and Display Options

The Propaq MD unit provides additional 12-Lead print and display options that you can specify
through the Supervisor parameter control panels (access to Supervisor is passcode-controlled).
Press the Setup quick access key (
select the four digits in the Supervisor passcode. Press SAVE when you are finished. Once you
have entered your supervisor passcode, you will be able to access the configurable options in
the Supervisor menu.
Select the Supervisor> ECG> 12-Lead option to display the 12-Lead parameter control panel:
9650-0806-01 Rev. D
Exit
)to restore the display of other monitoring functions.
12
Propaq MD Operator's Guide
12-Lead Print and Display Options
), then select Supervisor
.
Using the navigation keys,
14-13