Setting Upper And Lower Diastolic (Dia) Alarm Limits; Setting Upper And Lower Mean Arterial Pressure (Mean) Alarm Limits - ZOLL Propaq MD Operator's Manual

Defibrillator
Hide thumbs Also See for Propaq MD:
Table of Contents

Advertisement

Setting Upper and Lower Diastolic (DIA) Alarm Limits

Initially, the IBP Diastolic Alarm Settings menu specifies that the IBP diastolic pressure alarms
are enabled (
The upper and lower limits can be
default diastolic limits for adult, pediatric, and neonate patients, and gives the range in which
you can set these limits:
Patient Type
Adult
Pediatric
Neonate

Setting Upper and Lower Mean Arterial Pressure (MEAN) Alarm Limits

Initially, the IBP MEAN Alarm Settings menu specifies that IBP MEAN alarms are enabled
(
) or disabled (
ON
and lower limits can be
alarm limits for adult, pediatric, and neonate patients, and gives the range in which you can set
these limits:
Patient Type
Adult
Pediatric
Neonate
9650-0806-01 Rev. D
Enabling/Disabling IBP Alarms and Setting Alarm Limits
) or disabled (
)
,
and displays the default Upper and Lower diastolic limits.
ON
OFF
ON
IBP Diastolic Limit Default
Lower: 35 mmHg
Upper: 110 mmHg
Lower: 35 mmHg
Upper: 100 mmHg
Lower: 30 mmHg
Upper: 70 mmHg
)
,
and displays the default Upper and Lower MEAN limits. The upper
OFF
or
(default is
ON
OFF
IBP Mean Limit Default
Lower: 50 mmHg
Upper: 120 mmHg
Lower: 50 mmHg
Upper: 110 mmHg
Lower: 35 mmHg
Upper: 80 mmHg
Propaq MD Operator's Guide
or
(default is
). The following table lists the
OFF
OFF
IBP Diastolic Limit Range
Lower: -30 to 298 mmHg
Upper: -28 to 300 mmHg
Lower: -30 to 298 mmHg
Upper: -28 to 300 mmHg
Lower: -30 to 298 mmHg
Upper: -28 to 300 mmHg
). The following table lists the default MEAN
OFF
IBP Mean Limit Range
Lower: -30 to 298 mmHg
Upper: -28 to 300 mmHg
Lower: -30 to 298 mmHg
Upper: -28 to 300 mmHg
Lower: -30 to 298 mmHg
Upper: -28 to 300 mmHg
11-7