Defibrillator Testing With External Paddles - ZOLL Propaq MD Operator's Manual

Defibrillator
Hide thumbs Also See for Propaq MD:
Table of Contents

Advertisement

Defibrillator Testing with External Paddles

Prior to testing external defibrillator paddles with the Propaq MD unit, complete the testing
described in "Defibrillator/Pacing Test with Hands-Free Therapy Electrodes" on page 24-3.
Note:
Warning!
Keep hands, fingers, and other conductive materials away from paddle electrode plates
when discharging the defibrillator, or pacing.
Function
1
Press the Propaq MD power switch to turn the
unit on.
2
Connect the MFC to the defibrillator and
disconnect the patient end of the MFC from
any attached accessory (paddles or test
connector). Select Pads as the waveform
display source.
3
Connect the paddles set to the Propaq MD
MFC cable. Do not place the paddles
electrodes in contact with each other.
4
Press the RECORDER ON/OFF button on
the Sternum paddle.
5
Press the RECORDER ON/OFF button again. Verify that the recorder stops printing.
6
Press the Apex and Sternum paddle
electrodes together (face to face) maximizing
the contact area between the electrodes.
7
Press the ENERGY SELECT (+) button on
the Sternum paddle.
8
Press the ENERGY SELECT (+) button
again.
9
Repeatedly press the ENERGY SELECT (-)
button on the Sternum paddle until 30 Joules
is selected.
10 Press the CHARGE button on the Apex
paddle. (Press the CHARGE button twice if
the defibrillator window has disappeared.)
11 While the defibrillator is charged, press the
paddles electrodes together and then press
and hold the SHOCK button on the Apex
paddle (only).
9650-0806-01 Rev. D
If a low battery message appears during any of this testing, the battery is close to
depletion and should be replaced or recharged.
Propaq MD Operator's Guide
Defibrillator Testing with External Paddles
Response
Verify that the unit displays the SELF TEST
PASSED message.
Verify that a CHECK THERAPY
ELECTRODES message displays in the
waveform window
Verify that the CHECK THERAPY
ELECTRODES message changes to APPLY
PADDLES TO PATIENT.
Verify that the recorder begins printing.
Verify that the APPLY PADDLES TO PATIENT
message changes to SHORT DETECTED.
Verify that the defib window displays and the
following message appears:
PADDLE SHORT DETECTED
SELECT 30 J TO TEST
Verify that the selected energy increases to the
next highest level.
Verify that the defibrillator window shows
30 J as the selected energy.
Verify that a periodic tone sounds while the
defibrillator is charging. At the completion of
the charge cycle, the tone is continuous. The
charge time should be less than 7 seconds.
Verify that the CHARGE light on the Apex
handle illuminates when the defibrillator is
charged and ready to deliver energy.
Verify that the defibrillator does not
discharge.