IBP System Messages

The Propaq MD unit may display the following messages when monitoring IBP:
System Message
TRANSDUCER FAILURE
INCOMPATIBLE TRANSDUCER
CHECK PROBE
ZERO PROBE
ZEROING
IBP DISABLED
ZERO REJECTED
9650-0806-01 Rev. D
Cause
The IBP probe is damaged and needs to be replaced.
The IBP probe is not compatible. See the Appendix B,
Accessories, for a list of ZOLL-approved IBP probes.
The IBP probed has become disconnected.
The IBP probe is connected and needs to be zeroed.
The IBP probe is zeroing.
A system error has occurred, and the Propaq MD unit
should be serviced.
The IBP probe was not removed due to a pulsatile
pressure signal, excessive IBP artifact or excessive
transducer offset.
Propaq MD Operator's Guide
IBP System Messages