Defibrillator/Pacing Test With Hands-Free Therapy Electrodes - ZOLL Propaq MD Operator's Manual

Defibrillator
Hide thumbs Also See for Propaq MD:
Table of Contents

Advertisement

Defibrillator/Pacing Test with Hands-Free Therapy
Electrodes
Note:
Warning!
Keep hands, fingers, and other conductive materials away from paddle electrode plates
when discharging the defibrillator, or pacing.
Function
1
Connect the external AC adapter to a working
AC outlet and to the Propaq MD rear panel.
2
Insert a battery into the unit (if a battery is not
already in the slot).
3
Press the unit's power on switch.
4
Disconnect the external AC adapter from the
Propaq MD rear panel.
5
With no ECG cables attached to the unit,
select Lead I, II, or III as the waveform display
source.
6
Select Pads as the waveform display source
and connect the multifunction cable to the
unit. (Do not connect the test connector to the
MFC.)
7
Connect the multi-function cable (MFC) to the
test connector.
9650-0806-01 Rev. D

Defibrillator/Pacing Test with Hands-Free Therapy Electrodes

If a low battery message appears during any of this testing, the battery is close to
depletion and should be replaced or recharged.
Propaq MD Operator's Guide
Response
Verify that the green Auxiliary Power LED
illuminates on the Propaq MD front panel.
Verify that the Battery Charge LED on the
Propaq MD front panel illuminates yellow.
When the battery pack has been fully charged,
the battery status indicator light is green.
Verify the unit issues two audio beeps and that
the red, yellow, and green Visual Alarm
indicators illuminate for 2 to 3 seconds.
Note: The yellow LED may remain lit after the
unit powers on.
The Propaq MD unit performs a number of
power on self tests.
Verify that the unit displays SELF TEST
PASSED.
If the message SELF TEST FAILED appears
on the display, contact the appropriate
technical personnel or the ZOLL Technical
Service Department.
Verify that the unit continues to operate without
interruption on battery power and that the
displayed battery icon indicates that a fully
charged battery is installed.
Verify that no error messages are displayed.
Verify that a LEAD FAULT message displays in
the waveform window.
Verify that a CHECK THERAPY
ELECTRODES message displays in the
waveform window.
Verify that the CHECK THERAPY
ELECTRODES message is replaced by a
SHORT DETECTED message.
24-3