ZOLL Propaq MD Operator's Manual page 166

Defibrillator
Hide thumbs Also See for Propaq MD:
Table of Contents

Advertisement

C
13 A
HAPTER
UTOMATED
When a nonshockable rhythm is detected, the unit displays a NO SHOCK ADVISED message.
Immediately begin chest compressions and continue other treatment per protocol.
If the patient's rhythm is shockable, the unit displays the SHOCK ADVISED and PRESS
SHOCK messages. The defibrillator automatically prompts the operator to shock the patient at
the preconfigured energy level and the SHOCK button illuminates.
A continuous tone sounds for 20 or 50 seconds (depending on configuration), followed by a
higher pitch tone for 10 seconds.You must deliver the shock within this 30 or 60 second interval
(depending on configuration), or the defibrillator will disarm itself.
13–6
E
D
XTERNAL
EFIBRILLATOR
www.zoll.com
(AED) O
PERATION
NO
SHOCK
ADVISED
PRESS
SHOCK
9650-0806-01 Rev. D

Advertisement

Table of Contents
loading

Table of Contents