Enabling/Disabling Spo Alarms And Setting Alarm Limits - ZOLL Propaq MD Operator's Manual

Defibrillator
Hide thumbs Also See for Propaq MD:
Table of Contents

Advertisement

Enabling/Disabling SpO
When enabled, the Propaq MD unit sounds alarms whenever measurements are outside set
limits for the high and low SpO
You can enable (or disable) SpO
limits through the Alarms quick access key (
To configure SpO
1. Press the More quick access key (
Press
2.
3.
Press the Limits quick access key (
4.
Use the navigation buttons to highlight and select the appropriate alarm menu selection
(SpO
5.
On the selected Alarm Settings menu, use the navigation keys to select the fields that you
want to change. The fields are
Upper Limit Enable
Lower Limit Enable
Upper Limit
Lower Limit
When you are finished changing values on the alarm menu, navigate to the Backarrow key
6.
to exit the menu.
Setting Upper and Lower SpO
Initially, the SpO
disabled (
limits can be
adult, pediatric, and neonate patients, and gives the range in which you can set these limits.
Patient Type
Adult
Pediatric
Neonate
9650-0806-01 Rev. D
Enabling/Disabling SpO
Alarms and Setting Alarm Limits
2
values (and, if installed, SpCO and SpMet values).
2
2
SpCO, and SpMet alarms through the Alarms quick access key:
2
.
, SpCO, SpMet).
2
Alarm Limits
2
Alarm Settings menu specifies whether the SpO
2
),
and displays the default upper and lower SpO
OFF
or
(default is
ON
OFF
SpO
Limit Default
2
Lower: 85%
Upper: 100%
Lower: 85%
Upper: 100%
Lower: 85%
Upper: 95%
Propaq MD Operator's Guide
Alarms and Setting Alarm Limits
2
, SpCO, and SpMet alarms and set Upper and Lower alarm
)or through the SpO
).
).
Limits
).The following table lists the default SpO
OFF
SpO
Limit Range
2
Lower: 50 - 98%
Upper: 52 - 100%
Lower: 50 - 98%
Upper: 52 - 100%
Lower: 50 - 98%
Upper: 52 - 100%
Parameter Control Panel.
2
alarms are enabled (
2
limits. The upper and lower
2
limits for
2
)
or
ON
10-9