You can specify the following trace display ranges:
Display Type
Diagnostic
Filtered Diagnostic
Note:

Enabling 12-Lead Analysis

This option allows you to enable or disable 12 Lead Analysis. The default is enabled (
9650-0806-01 Rev. D
Frequency Response
0.525 to 150 Hz
0.525 to 40 Hz
The display and the recording present the waveform according to the filter setting
(Diagnostic or Filtered Diagnostic), but the 12-Lead Interpretive Analysis is always
performed using a diagnostic bandwidth that follows the unit's AC Mains filter setting.
Propaq MD Operator's Guide
12-Lead Print and Display Options
On
).
14-15