If you don't select a label, the Temperature channels appear with the default labels of T1 and
T2.

Temperature System Messages

The Propaq MD unit may display the following messages when monitoring Temperature.
Note:
System Message
CHECK PROBE
PROBE FAILURE
TEMP DISABLED
9650-0806-01 Rev. D
Description
Arterial temperature probe
Central body temperature or tympanic membrane temperature
probe
Esophageal temperature probe
Rectal temperature probe
Skin temperature probe (surface application)
Ventilator airway temperature probe
Nasopharynx or nasal/oral temperature probe
The temperature function performs a self test when initially powered on and also
performs system tests automatically, every 10 seconds, while this function is active.
Propaq MD Operator's Guide
Temperature System Messages
Cause
The temperature probe is disconnected. Check probe
and reconnect it.
The temperature probe is defective. Replace the
Temperature probe.
A system error has occurred. The Propaq MD unit cannot
take temperature measurements and should be serviced.