Printer Setup
To configure the printer functions, press the More quick access key (
access key (
desired print configuration. The Setup>Printer window contains the following print
configuration options:
Print configuration
Print Speed
Print Number of Traces
Print Grid
Print on Snapshot
Print on Patient Alarm
Print on NIBP
Print on Trend
Automatic Prints
You can print any following events by setting the monitor to print automatically in the
Setup>Supervisor>Printer window. To do this, press (
navigation buttons, select Supervisor, and then Printer.
Defib/Pacer Events
•
Event
•
Treatment Snapshot
•
Treatment Summary Snapshots
•
Printing Waveforms
You can print waveforms by pressing
waveforms as configured in the Number of Traces option. The Print Number of Traces option
can be set to
Note:
The ECG Waveform
ECG waveforms are printed on a grid with major divisions (dotted lines) every 5mm and minor
divisions (single dots) every 1 mm. The ECG waveform is always printed if ECG is monitored.
The Invasive Pressure Waveform
Pressure waveforms are printed on a grid with major divisions (vertical dotted lines) every 5
mm, and the pressure scale grids are printed horizontally.
23–2
), highlight Printer, and press Select. Use the navigation keys to select the
,
, or
. Patient vital sign numerics are printed above the waveforms.
1
2
3
Waveforms that have invalid data print a dashed line for that trace.
www.zoll.com
Options
25 mm or 50 mm per second
1, 2, 3, or 4
On or Off
On or Off
On or Off
On or Off
On or Off
), then press
. The Propaq MD unit will print the displayed
), press the Setup quick
. Using the
9650-0806-01 Rev. D