Date: ________________
Mfr/Model No.: ____________________________ Serial No. or Facility ID No.: _________________
At the beginning of each shift, inspect the unit. Indicate whether all requirements have been met.
1. Defibrillator Unit
Clean, no spills, and casing undamaged
2. Cables/Connectors
a. Inspect for damaged insulation, frayed/broken wires, or bent
connector pins
b. Connectors engage securely
3. Sensors (pulse oximetry, NIBP cuff and hose, temperature sensors)
a. Inspect for signs of damage or excessive wear
b. Connectors engage securely
4. Paddles
a. Clean, not pitted, or damaged.
b. Switches operate freely
c. Cables and connectors free of damage and engage securely
5. Supplies and Accessories
a. Therapy pads in sealed pouches (2 sets, not expired)
b. Defib gel or gel pads
c. ECG monitoring electrodes
d. Alcohol wipes
e. Razors/scissors
f. Recorder paper
6. Batteries
a. Fully charged battery installed in unit
b. Fully charged spare battery available
7.
Ready For Use Indicator shows Ready - No flashing or Do Not Use symbol
8. Operational Checks
A. Power Up Sequence
a. Battery charge and AC indicators illuminate with AC power
b. Audio beeps heard and Visual Alarm indicators briefly illuminate
c. Self Test passed
B. Hands Free Defibrillation (Test with only battery power)
a. CHECK THERAPY ELECTRODES and PADS SHORT
DETECTED messages display
b. Charge time < 7 seconds
c. DEFIB SHORT TEST passed at 30 Joules
C. Pacer Test (Test with only battery power)
a. Printer prints Pace markers every 25 mm at 60 ppm
b. PACING: SHORT DETECTED message displays-- no error
messages at 100 mA
c. PACING: CHECK THERAPY ELECTRODES message displays
D. Paddles
a. APPLY PADDLES TO PATIENT message displays when paddles
are connected to MFC
b. Paddle switches functional (Recorder, Energy Select, Charge, Shock)
c. DEFIB SHORT TEST passed at 30 Joules
E. Reconnect Unit to AC Power
____________ Major problem(s) identified
Signature
9650-0806-01 Rev. D
Propaq MD OPERATOR'S SHIFT CHECKLIST
Shift: _______________
Note any corrective actions taken. Sign the form.
(OUT OF SERVICE)
_______________________________________________
Propaq MD Operator's Guide
Location: _______________
Okay as found
Corrective Action/Remarks
24-11