Cleaning And Reuse Of Sensors; Connecting The Spo Sensor - ZOLL Propaq MD Operator's Manual

Defibrillator
Hide thumbs Also See for Propaq MD:
Table of Contents

Advertisement

C
10 P
CO-O
HAPTER
ULSE

Cleaning and Reuse of Sensors

Reusable sensors can be cleaned as follows:
Disconnect the sensor from the patient cable, if appropriate. Wipe the entire sensor clean with a
70% isopropyl alcohol moistened pad. Allow the sensor to air dry before returning it to use.
Connecting the SpO
To connect the SpO
1. When using a sensor extension cable, inspect the cable before use. Replace the cable if it
shows any signs of wear, breakage, or fraying. Plug the sensor extension cable into the SpO
receptacle on the side of the Propaq MD unit:
Figure 10-3 Connecting the SpO
Displaying SpO
When the connection is made between the SpO
INITIALIZING, appears in the SpO
measurement. If SpCO and SpMet are installed, the SpCO and SpMet measurements will
alternate every two seconds underneath the SpO
SpO
2
Measurement
If the message, SENSOR FAILURE, appears, the sensor is either incompatible with the Propaq
MD unit, or it is not working, and you will need to replace the sensor.
A blip bar appears on the right side of the SpO
the pulse derived from the plethysmograph normalized waveform.
10-8
(S
O2, SPCO,
XIMETRY
P
Sensor
2
sensor to the Propaq MD unit:
2
Sensor to the Propaq MD Unit
2
, SpCO, and SpMet Measurements
2
display. After a brief delay, the unit displays the SpO
2
%
SpO2
Blip bar
97
Alternating SpMet
SpMet 1.3
Measurement
www.zoll.com
S
M
)
AND
P
ET
sensor and the unit, the message,
2
display.
2
%
SpO2
97
SpCO 2
numeric display window. The blip bar indicates
2
2
Blip bar
Alternating SpCO
Measurement
9650-0806-01 Rev. D