AC Power Outlet for Peripheral

The REAR ASSY and REAR CONN 220V ASSY are the rear-panel assemblies which interconnect
external signals and power with the LOGIQ 5 system. Both of the assemblies are the same except the
circuit breaker's current ratings depending on voltage of the system power source (See Safety
described below).
5-8-4-1
Safety
Circuit Breaker
Table 5-11
Assembly
REAR CONN
REAR CONN 220V
Electrical Isolation
The rear-panel's Insite (modem) interface is electrically isolated from the system's internal circuit.
The isolator's dielectric strength should be 500VAC RMS or higher within 1 minute (in compliance
with IEC60601-1-1).
5-8-4-2
Camera (Printer) Remote Control Interface
The Rear-panel has two remote control interface terminals for B/W and color printers.
Table 5-12
Color camera (shutter) control
B/W camera (printer) control
7
Circuit Breaker Specification
Power Voltage Rating
(V)
110
220
Camera (Printer) Remote Control Interface Specifications
Chapter 5 Components and Functions (Theory)
Peripheral Current
Trip Time
Rating (A)
1 hour max. @ 5.4A, 25
5.0
degreeC
1 hour max. @ 3.5A, 25
3.0
degreeC
Level (V)
Active Polarity
H:2.0/L:0.8
H:2.0/L:1.0
LOGIQ™5 PRO S
ERVICE
Reset Time
60 seconds
180 seconds
Retention Time
L
60 msec
L
100 msec
M