Basic Measurements

NOTE:
The following instructions assume that you first scan the patient and then press Freeze.
4-3-6-1
Distance and Tissue Depth Measurements
1.) Press MEASURE once; an active caliper displays.
2.) To position the active caliper at the start point (distance) or the most anterior point (tissue
3.) To fix the start point, press Set. The system fixes the first caliper and displays a second active
4.) To position the second active caliper at the end point (distance) or the most posterior point
5.) To complete the measurement, press SET. The system displays the distance or tissue depth
Before you complete a measurement:
To toggle between active calipers, press MEASURE.
To erase the second caliper and the current data measured and start the measurement again, press
CLEAR once.
NOTE:
To rotate through and activate previously fixed calipers, turn CURSOR SELECT.
NOTE:
After you complete the measurement, to erase all data that has been measured to this point,
but not data entered onto worksheets, press CLEAR.
4-3-6-2
Circumference/Area (Ellipse) Measurement
1.) Press MEASURE once; an active caliper displays.
2.) To position the active caliper, move the TRACKBALL.
3.) To fix the start point, press SET. The system fixes the first caliper and displays a second active
4.) To position the second caliper, move the TRACKBALL.
5.) Turn the ELLIPSE control; an ellipse with an initial circle shape appears.
NOTE:
Be careful not to press the Ellipse control as this activates the Body Pattern.
6.) To position the ellipse and to size the measured axes (move the calipers), move the
7.) To increase the size, turn the ELLIPSE control in a clockwise direction. To decrease the size,
8.) To toggle between active calipers, press MEASURE.
9.) To complete the measurement, press SET. The system displays the circumference and area
Before you complete a measurement:
-
-
4 - 10
7
depth), move the TRACKBALL.
caliper.
(tissue depth), move the TRACKBALL.
value in the measurement results window.
caliper.
TRACKBALL.
turn the ELLIPSE control in a counterclockwise direction.
in the measurement results window.
To erase the ellipse and the current data measured, press CLEAR once. The original caliper
is displayed to restart the measurement.
To exit the measurement function without completing the measurement, press CLEAR a
second time.
Section 4-3 - General Procedure
LOGIQ™ 5 PRO S
ERVICE
M
ANUAL