LMT Assy (FRU No. 501)

Purpose: This is a description on how to remove and replace the LMT Assy.
8-7-2-1
Tools
Common pillips screwdrivers
8-7-2-2
Needed Manpower
1 person, 25 minutes + travel
8-7-2-3
Preparations
Shut Down the System and switch off the Main Breaker at the rear as described in section 4-3-2 on
page 4.
8-7-2-4
Removal Procedure
An electronic discharge may damage a component. Turn OFF power and wear the wrist strap
CAUTION
before you remove circuit boards. Do not unplug the power cord to keep ground continuity.
Do not bend or flex the boards when mounting/dismounting each boards. Surface mount IC
boards are very susceptible to damage from flex/torque.
1.) Remove the RLY Assy. Refer to section 8-7-1 on page 88.
2.) Unscrew two (2) screws (1-2) to remove the LMT Board. Refer to Figure 8-82.
7
Figure 8-82 Unscrew 2 screws to remove the LMT B'd
Chapter 8 Replacement Procedures
LOGIQ™ 5 PRO S
1
1
2
2
M
ERVICE
ANUAL
8 - 91