Rly Block Diagram - GE LOGIQ 5 PRO Service Manual

Table of Contents

Advertisement

GE M
S
EDICAL
YSTEMS
D
2380207, R
IRECTION
EVISION
5-4-1
RLY
128
Probe A
128
Probe B
128
Probe C
DCWD
Probe I/F
RLY ASSY contains of 3 phased array probe connectors and provides switchable connection between
probes and 64ch transmitters/receivers.
NOTE:
There is no numeric order in the probe port, system recognizes the first probe slotted on the
port as the probe #1, regardless of the probe port location.
This board also has interface for a single CWD probe connector.
The main function of RLY ASSY is as follows.
-
-
-
-
-
-
-
-
5-6
7
RELAY
Probe ID
I/F
Figure 5-5 RLY Block Diagram
3-to-1 selectors for three probes.
Interface with JUSC bus (control bus)
Interface with FEC ASSY for IIC bus.
Interface with DCWD Connector.
Supply/Cut control and failure detection of supply voltage for Mux circuit in a Probe.
Device: Mechanical Relay
Switch: 128ch 3:1 selector
Temperature sensor in this assy detects temperature of RLY assy.
Section 5-4 - Front End
RELAY
128
128
Control
Circuit
JUSC I/F
LOGIQ™5 PRO S
ERVICE
64
MUX
To LMT
SHV
Control
Temperature
FPGA
Sensor
M
ANUAL

Hide quick links:

Advertisement