Before Starting Software Loading

8-10-5-1
Check Software Version and BEP Type
1.) Turn ON the system
2.) Press Utility button on the Front Panel.
NOTE:
Operator Login Window appears. Select adm(Administrator) for the Operator field.
Adm(Administrator) will be shown in the Operator field as default. Enter password which is
configured already ( or the password might not be necessary). Then click on Log on.
3.) Touch the Utility button on the FrontPanel
4.) Click System and then About tab.
5.) Check Image Part Number is 2380104 or 5134560 and 5131757 or 5132192 then confirm BEP
type. Refer to Figure 8-38 and Table 8-3
Table 8-56 BEP Type
2380104-x or 5134559-x
5131757-x or 5132191-x
Software version & Part number
8 - 168
7
Figure 8-187 Operator Login Window
Image Part Number
Figure 8-188 Check BEP Type and Software Version
Section 8-10 - Software Loading Procedure
LOGIQ™ 5 PRO S
BEP Type
BEP2