Op Panel Trackball Assy (Fru No. 208) - GE LOGIQ 5 PRO Service Manual

Table of Contents

Advertisement

GE M
S
EDICAL
YSTEMS
D
2380207, R
IRECTION
EVISION
8-4-9

OP Panel Trackball Assy (FRU No. 208)

Purpose: This is a description on how to remove and replace OP Panel Trackball Assy.
8-4-9-1
Tools
Common pillips screwdrivers.
8-4-9-2
Needed Manpower
1person, 5 minutes + travel
8-4-9-3
Preparations
Shut Down the System and switch off the Main Breaker at the rear as described in section 4-3-2 on
page 4.
Keyboard Assy should be removed before proceed. For more information, refer to section 8-4-1 on
page 16.
8-4-9-4
Removal Procedure
1.) Disconnect two connectors (1-2) from the Trackball Assy. Refer to Figure 8-34.
2.) Unscrew three screws (3-5) to remove the Trackball Assy. Refer to Figure 8-35.
7
2
Figure 8-34 Disconnect two connectors
4
5
Figure 8-35 Unscrew 3 screws
Chapter 8 Replacement Procedures
LOGIQ™ 5 PRO S
1
3
M
ERVICE
ANUAL
8 - 33