Recommended And Alternate Ultrasound Room Layout - GE LOGIQ 5 PRO Service Manual

Table of Contents

Advertisement

GE M
S
EDICAL
YSTEMS
D
2380207, R
IRECTION
EVISION
2-3-4

Recommended and Alternate Ultrasound Room Layout

Recommended standard floor plan and a minimal floor plan for ultrasound equipment:
FILM VIEWER
FILM SUPPLIES
DOOR
30 IN.
(76 CM)
7
FILM VIEWER
COUNTER TOP
SECRETARYS OR
DOCTOR'S DESK
A 14 by 17 foot Recommended Floor Plan
LINEN SUPPLY
SINK
PROBES/SUPPLIES
EXTERNAL
PERIPHERALS
FOOT
LOGIQ 5
SW
CONSOLE
STOOL
EXAMINARION
TABLE
24 IN.
76 IN.
(61 CM)
(193 CM)
An 8 by 10 foot Minimal Floor Plan
Figure 2-1 RECOMMENDED ULTRASOUND ROOM LAYOUT
Chapter 2 Pre-Installation
LOGIQ™5 PRO S
DEDICATED ALALOG TELEPHONE
LINE FOR CONNECTION TO INSITE
18 IN.
(46 CM)
DEDICATED
POWER
RECEPTACLE
CONSOLE
FOOT
SW
STOOL
SINK
76 IN.
(193 CM)
SUCTION LINE
EMERGECY
OXYGEN
OVERHEAD
LIGHTS DIMMER
DOOR
42 IN.
(107 CM)
Scale : Each square equals one square foot
DEDICATED POWER
OUTLETS
DEDICATED ANALOG TELEPHONE
LINE FOR CONNECTION TO INSITE
GE CABINET
FOT SOFTWARE
AND MANUALS
M
ERVICE
ANUAL
LINEN SUPPLY
24 IN.
(61 CM)
PATIENT
TOILET
FACILITY
2-9

Hide quick links:

Advertisement

Table of Contents
loading