GE M
S
EDICAL
YSTEMS
D
2380207, R
IRECTION
EVISION
8-9-3
JPC Assy (FRU No. 702)
Purpose: This is a description on how to remove and replace the JPC Assy.
8-9-3-1
Tools
•
Common pillips screwdrivers
•
Allen/Unbraco wrench
8-9-3-2
Needed Manpower
•
1 person, 15 minutes + travel
8-9-3-3
Preparations
•
Shut Down the System and switch off the Main Breaker at the rear as described in section 4-3-2 on
page 4.
8-9-3-4
Removal Procedure
Do not wear the ESD wrist strap when you remove a part of power supply unit. Turn OFF power
CAUTION
and unplug the power cord before removing a part of power supply unit. However be sure to turn
off power and wear the strap before you remove a circuit boards.
1.) Unscrew nine (9) hexagon screws (1-9) to remove the AC Power Top cover.
2.) Unscrew five (5) screws (10-14).
JPC Assy
8 - 158
7
Figure 8-176 JPC Assy
Section 8-9 - Power Block
LOGIQ™ 5 PRO S
ERVICE
M