Leveled Sine Wave Specifications - Fluke 5520A Service Manual

Multi-product calibrator
Hide thumbs Also See for 5520A:
Table of Contents

Advertisement

5520A
Service Manual
6-6.

Leveled Sine Wave Specifications

Leveled Sine Wave
Characteristics
into 50
Amplitude Characteristics (for measuring oscilloscope bandwidth)
Range (p-p)
Resolution
Adjustment Range
1-Year Absolute
Uncertainty,
tcal
5 C
Flatness (relative to
50 kHz)
Short-Term Amplitude
Stability
Frequency Characteristics
Resolution
1-Year Absolute
Uncertainty,
tcal
5 C
Distortion Characteristics
2nd Harmonic
3rd and Higher
Harmonics
[1] Within one hour after reference amplitude setting, provided temperature varies no more than
6-8
Table 6-3. Leveled Sine Wave Specifications
50 kHz
50 kHz to
(reference)
100 MHz
(2% of
(3.5% of
output
output
+ 300 V)
+ 300 V)
not applicable
(1.5% of
output
+ 100 V)
Frequency Range
100 MHz to
300 MHz
5 mV to 5.5 V
< 100 mV: 3 digits
100 mV: 4 digits
continuously adjustable
(4% of output
+ 300 V)
(2% of
output
+ 100 V)
1% [1]
10 kHz
2.5 ppm
-33 dBc
-38 dBc
300 MHz to
600 MHz
(6% of output
+ 300 V)
(4% of output
+ 100 V)
5°C.