Fluke 5520A Service Manual page 88

Multi-product calibrator
Hide thumbs Also See for 5520A:
Table of Contents

Advertisement

5520A
Service Manual
EOFSTR?
Description: Returns the End-Of-File character string used for calibration reports
Parameter:
Response:
PR_RPT
Description: Prints a self-calibration report out the selected serial port
Parameter:
Example:
RPT?
Description: Returns a self-calibration report.
Parameter:
Example:
RPT_PLEN
Description: Sets the page length used for calibration reports. This setting is stored in
Parameter:
RPT_PLEN?
Description: Returns the page length used for calibration reports.
Parameter:
Response:
3-32
None
(String) The End-Of-File character string
1. Type of report to print: STORED, ACTIVE, or CONSTS
2. Format of report: PRINT (designed to be read)
SPREAD (designed to be loaded into a spreadsheet)
3. Calibration interval to be used for instrument specifications in the
report: I90D (90 day specifications) or I1Y (1 year specifications)
4. Serial port out which to print report: HOST or UUT
PR_RPT STORED,PRINT,I90D,HOST
1. Type of report to return: STORED, ACTIVE, or CONSTS
2. Format of report: PRINT (designed to be read)
SPREAD (designed to be loaded into a spreadsheet)
3. Calibration interval to be used for instrument specifications in the
report:
I90D (90 day specifications) or I1Y (1 year specifications)
RPT? STORED,PRINT,I90D
nonvolatile memory.
Page length
None
(Integer) Page length