3-26. Verifying Capacitance

Verify that the 5520A performance is within the limits in Table 3-28. Use the PM 6304C
RCL Meter directly for capacitance values up to and including 109.000 F. Above
109.000 F, you must use a timed charge up routine with a constant current source in
order to achieve the required test uncertainty ratio.
To verify capacitance greater than 109.000 F, see the section titled "200 F to 110 mF
Capacitance Verification" found later in this chapter.
Range
0.3999 nF
0.3999 nF
1.0999 nF
1.0999 nF
1.0999 nF
3.2999 nF
10.9999 nF
10.9999 nF
32.9999 nF
109.999 nF
109.999 nF
329.999 nF
329.999 nF
3-46
Table 3-27. Verification Tests for AC Current (cont)
Output
Frequency
10.9000 A
45 Hz
10.9000 A
65 Hz
10.9000 A
500 Hz
10.9000 A
1 kHz
10.9000 A
5 kHz
20.0000 A
45 Hz
20.0000 A
65 Hz
20.0000 A
500 Hz
20.0000 A
1 kHz
20.0000 A
5 kHz
Table 3-28. Verification Tests for Capacitance
Test Frequency or
Output
Current
0.1900 nF
5 kHz
0.3500 nF
1 kHz
0.4800 nF
1 kHz
0.6000 nF
1 kHz
1.0000 nF
1 kHz
2.0000 nF
1 kHz
7.0000 nF
1 kHz
10.9000 nF
1 kHz
20.0000 nF
1 kHz
70.000 nF
1 kHz
109.000 nF
1 kHz
200.000 nF
1 kHz
300.000 nF
1 kHz
Lower Limit
Upper Limit
10.8926 A
10.9075 A
10.8926 A
10.9075 A
10.8893 A
10.9107 A
10.8893 A
10.9107 A
10.6255 A
11.1745 A
19.9750 A
20.0250 A
19.9750 A
20.0250 A
19.9690 A
20.0310 A
19.9690 A
20.0310 A
19.4950 A
20.5050 A
Lower Limit
0.1793 nF
0.3387 nF
0.4682 nF
0.5877 nF
0.9862 nF
1.9824 nF
6.9767 nF
10.8693 nF
19.8620 nF
69.767 nF
108.693 nF
199.320 nF
299.130 nF
Upper Limit
0.2007 nF
0.3613 nF
0.4918 nF
0.6123 nF
1.0138 nF
2.0176 nF
7.0233 nF
10.9307 nF
20.1380 nF
70.233 nF
109.307 nF
200.680 nF
300.870 nF