14. Measure and enter the values into the UUT for calibration steps 14 and 15 in Table 3-
15. Disconnect the DMM from the UUT, and connect it to the Guildline 9334H/1G
16. Connect the UUT to the DMM in a 2-wire ohms configuration as shown in
17. Measure and enter the value into the UUT for calibration step 16 in Table 3-15 as
18. Verify that the UUT is in Standby and disconnect the test equipment.
Step
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
3-22
15 as prompted.
Resistance Standard. Scale the 1 G DMM range to the Resistance Standard as
described in the HP3458A user documentation.
Figure 3-12.
prompted.
Table 3-15. Calibration Steps for Resistance
5520A Output (4-Wire Ohms, NORMAL and AUX)
1.0000
11.0000
110.0000
0.350000 k
1.100000 k
3.50000 k
11.00000 k
35.0000 k
2-Wire Ohms, NORMAL
110.0000 k
0.350000 M
1.100000 M
3.50000 M
11.00000 M
35.0000 M
110.000 M
400.00 M