3. Enter 10 MHz into the Calibrator Mainframe. Allow the power meter reading to
4. Enter the next frequency listed in Table 6-56. Allow the power meter's reading to
5. Enter 10 MHz into the Calibrator Mainframe. Allow the power meter reading to
6. Repeat steps 4 and 5 for all of frequencies listed in Table 6-56. Continue until you
7. When you have completed Columns A and B, press
Calibrator
Mainframe
Freq. (MHz)
20
50
100
125
160
200
220
235
250
300
Complete Columns A-G as follows:
A
B
C
D
E
F
G
6-102
stabilize, then enter the power meter's reading in Column B of Table 6-56.
stabilize, then enter the reading into Column A of the table.
stabilize, then enter the power meter's reading in Column B of Table 6-56.
have completed Columns A and B.
Mainframe's output. Complete Table 6-56 by performing the calculations for each
column. Compare Column G to the specifications listed in the final column.
Table 6-56. High Frequency Flatness Verification at 5.5 V
B
A
10 MHz
Enter the 437B present frequency Reading (W).
Enter the 437B 10 MHz Reading (W).
Apply power sensor correction factor for present frequency (W): CF * (Column A entry)
Apply power sensor correction factor for 10 MHz (W): CF * (Column B entry)
Compute and enter Error relative to 10 MHz (%): 100 * (sqrt(Column A entry) -
sqrt(Column B entry)) / sqrt(Column B entry)
Enter the 10 MHz rms Error (%) for 5.5 V from Table 6-55, Column C.
Compute and enter the Calibrator Mainframe Flatness Deviation (%): (Column E entry)
+ (Column F entry)
C
D
E
to remove the Calibrator
Calibrator
F
G
Mainframe
Flatness Spec. (%)
1.50 +100 uV
1.50 +100 uV
1.50 +100 uV
2.00 + 100 uV
2.00 + 100 uV
2.00 + 100 uV
2.00 + 100 uV
2.00 + 100 uV
2.00 + 100 uV
2.00 + 100 uV