Fluke 5520A Service Manual page 185

Multi-product calibrator
Hide thumbs Also See for 5520A:
Table of Contents

Advertisement

Connect the HP 437B Power Meter to either the 8482A or the 8481D Power Sensor as
shown in Figure 6-11. For more information on connecting the two instruments, see the
power meter and power sensor operators manuals.
Connect the power meter/power sensor combination to the SCOPE connector on the
Calibrator Mainframe, as shown in Figure 6-12.
The Hewlett-Packard 437B Power Meter must be configured by setting the parameters
listed below. Zero and self-calibrate the power meter with the power sensor being used.
Refer to the Hewlett-Packard 437B operators manual for details.
PRESET
RESOLN 3
AUTO FILTER
WATTS
SENSOR TABLE 0 (default)
Figure 6-11. Connecting the HP 437B Power Meter to the HP 8482A or 8481D Power Sensor
Figure 6-12. Connecting the Calibrator Mainframe to the HP Power Meter and Power Sensor
5520A
CALIBRATOR
NORMAL
AUX
SCOPE
V, , ,RTD
A, -SENSE, AUX V
OUT
PREV
STBY
OPR
EARTH
SCOPE
BOOST
MENU
HI
1000V
20V
1V PK
RMS
RMS
MAX
dBm
sec
MAX
MAX
7
8
9
m
V
Hz
SETUP
LO
150V
PK
MAX
TRIG
20V
n
W
RMS
¡F
NEW
4
5
6
MAX
k
A
¡C
REF
p
MEAS
20V
1
2
3
M
F
SHELLS
PK
TC
GUARD
20A
NOT
MAX
GROUNDED
MULT
/ +
0
SHIFT
ENTER
x
20V PK MAX
TC
20V PK MAX
SC600 Option
Verification
OM035f.eps
EDIT
RESET
FIELD
CE
TRIG
OUT
POWER
DIV
I
O
yg036f.eps
6
6-45

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents