5520A
Service Manual
Calibrator Mainframe output
0 mV
1.25 mV
-1.25 mV
2.49 mV
-2.49 mV
2.5 mV
-2.5 mV
6.25 mV
-6.25 mV
9.90 mV
-9.90 mV
10.0 mV
-10.0 mV
17.5 mV
-17.5 mV
24.9 mV
-24.9 mV
25.0 mV
-25.0 mV
67.5 mV
-67.5 mV
109.9 mV
-109.9 mV
110 mV
-110 mV
305 mV
-305 mV
499 mV
-499 mV
0.50 V
-0.50 V
1.35 V
-1.35 V
2.19 V
-2.19 V
2.20 V
-2.20 V
6.60 V
-6.60 V
10.99 V
-10.99 V
11.0 V
-11.0 V
70.5 V
-70.5 V
130.0 V
-130.0 V
6-30
Table 6-19. DC Voltage Verification at 1 M
HP 3458A Reading (V DC)
Tolerance (V DC)
0.00004 V
4.063E-05 V
4.063E-05 V
4.125E-05 V
4.125E-05 V
4.125E-05 V
4.125E-05 V
4.313E-05 V
4.313E-05 V
4.495E-05 V
4.495E-05 V
0.000045 V
0.000045 V
4.875E-05 V
4.875E-05 V
5.245E-05 V
5.245E-05 V
0.0000525 V
0.0000525 V
7.375E-05 V
7.375E-05 V
9.495E-05 V
9.495E-05 V
0.000095 V
0.000095 V
0.0001925 V
0.0001925 V
0.0002895 V
0.0002895 V
0.00029 V
0.00029 V
0.000715 V
0.000715 V
0.001135 V
0.001135 V
0.00114 V
0.00114 V
0.00334 V
0.00334 V
0.005535 V
0.005535 V
0.00554 V
0.00554 V
0.03529 V
0.03529 V
0.06504 V
0.06504 V