11. Set the oscilloscope vertical to 5 mV/div. Check the aberrations.
12. Check for rise time < 300 ps at 250 mV, 1 V, and 2.5 V outputs.
1st Aberration
2nd Aberration
3rd Aberration
R36
R12
R13
T
R35
om050f.eps
Figure 6-17. Adjusting Edge Aberrations