Leveled Sine Wave Specifications - Fluke 5500A Service Manual

Multi-product calibrator
Hide thumbs Also See for 5500A:
Table of Contents

Advertisement

5500A
Service Manual
6-6.

Leveled Sine Wave Specifications

Leveled Sine Wave
Characteristics
into 50 Ω
Amplitude Characteristics (for measuring oscilloscope bandwidth)
Range (p-p)
Resolution
Adjustment Range
1-Year Absolute
Uncertainty,
tcal ± 5 °C
Flatness (relative to
50 kHz)
Short-Term Amplitude
Stability
Frequency Characteristics
Resolution
1-Year Absolute
Uncertainty,
tcal ± 5 °C
Distortion Characteristics
2nd Harmonic
3rd and Higher
Harmonics
[1] Within one hour after reference amplitude setting, provided temperature varies no more than ± 5°C.
6-8
Table 6-3. Leveled Sine Wave Specifications
50 kHz
50 kHz to
(reference)
100 MHz
± (2% of
± (3.5% of
output
output
+ 300 µV)
+ 300 µV)
not applicable
± (1.5% of
output
+ 100 µV)
Frequency Range
100 MHz to
300 MHz
5 mV to 5.5 V
< 100 mV: 3 digits
≥ 100 mV: 4 digits
continuously adjustable
± (4% of output
+ 300 µV)
± (2% of output
+ 100 µV)
≤ 1% [1]
10 kHz
± 2.5 ppm
≤ -33 dBc
≤ -38 dBc
300 MHz to
600 MHz
± (6% of output
+ 300 µV)
± (4% of output
+ 100 µV)