Introduction; Equipment Required For Calibration And Verification - Fluke 5520A Service Manual

Multi-product calibrator
Hide thumbs Also See for 5520A:
Table of Contents

Advertisement

3-1. Introduction

You should recalibrate at the end of either a 90-day or 1-year calibration interval. If you
recalibrate every 90 days, use the 90-day specifications, which provide higher
performance. Use the Verification procedure or any part thereof any time you need to
verify that the Calibrator is meeting its specifications.
Fluke recommends that you return the 5520A to Fluke for calibration and verification.
The Fluke Service Center uses a software-controlled verification process and provides a
detailed test report including traceability to national standards. If you plan to calibrate or
verify the 5520A at your site, use this chapter as a guide. The procedures in this chapter
are manual versions of the software-controlled process used at the Fluke Service Center.

3-2. Equipment Required for Calibration and Verification

The equipment listed in Table 3-1 is required to calibrate and verify performance of the
5520A. If a specified instrument is not available, you can substitute an instrument that
has the same, or better performance.
Table 3-1. Consolidated List of Required Equipment for Calibration and Verification
Quan.
Manufacturer
1
Fluke
1
Hewlett
Packard
1
Fluke
1
Keithley
1
Fluke
1
Fluke
1
Fluke
1
Fluke
1
Guildline
1
Guildline
1
Fluke
1
Fluke
1
Guildline
1
Guildline
1
Fluke
1
Fluke
Model
Equipment
5500A/LEADS Test lead set
3458A
DMM
with -002
Option
752A
Reference Divider 100:1, 10:1
155
Null Detector
742A-1k
Resistance Standard, 1 k
742A-100
Resistance Standard, 100
742A-10
Resistance Standard, 10
742A-1
Resistance Standard, 1
9230
0.1
shunt
9230
0.01
shunt
742A-1M
Resistance Standard, 1 M
742A-10 M
Resistance Standard, 10 M
9334/100 M
Resistance Standard, 100 M
9334/1G
Resistance Standard, 1G
PN 900394
Type N to dual banana adapter AC voltage
5790A
AC Measurement Standard
Calibration and Verification

Introduction

Purpose
All functions
DC voltage, dc current,
resistance, capacitance,
thermocouple measurement
and sourcing
DC voltage
DC voltage (calibrate Fluke
752A for dc voltage)
DC current
DC current
DC current
DC current
DC current, verification
procedure only
DC current
Resistance
Resistance
Resistance
Resistance
AC voltage, ac current
3
3-3

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents