Sc300 Hardware Adjustments; Equipment Required - Fluke 5520A Service Manual

Multi-product calibrator
Hide thumbs Also See for 5520A:
Table of Contents

Advertisement

Calibrator
Calibrator
Mainframe
Mainframe
output
Wave Type
(@ 10 kHz)
square
5.0 mV
square
10.9 mV
square
45 mV
square
109 mV
square
0.45V
square
1.09V
square
2.20V
sine
5.0 mV
sine
10.9 mV
sine
45 mV
sine
109 mV
sine
0.45 V
sine
1.09 V
sine
2.20 V
triangle
5.0 mV
triangle
10.9 mV
triangle
45 mV
triangle
109 mV
triangle
0.45 V
triangle
1.09 V
triangle
2.20 V

6-136. SC300 Hardware Adjustments

Hardware adjustments must be made to the leveled sine and edge functions each time the
SC300 is repaired. In addition to the adjustment procedures, this section provides lists of
the required equipment and some recommendations on models that have the capabilities
required by these procedures. Equivalent models can be substituted if necessary.

6-137. Equipment Required

The following equipment is necessary for performing the hardware adjustments
described in this section. The models listed are recommended for providing accurate
results.
Standard adjustment tool for adjusting the pots and trimmer caps
Extender Card (pn 661865, 5800A-7006K, Extender Kit )
Table 6-59. Wave Generator Verification at 50
5790A
Conversion
Reading
Factor
(V rms)
2.0000
2.0000
2.0000
2.0000
2.0000
2.0000
2.0000
2.8284
2.8284
2.8284
2.8284
2.8284
2.8284
2.8284
3.4641
3.4641
3.4641
3.4641
3.4641
3.4641
3.4641

SC300 Hardware Adjustments

5790A Reading x
Conversion Factor
(V p-p)
250.00 V
430.00 V
1.450 mV
3.370 mV
13.570 mV
32.500 mV
66.100 mV
250.00 V
430.00 V
1.450 mV
3.370 mV
13.570 mV
32.500 mV
66.100 mV
250.00 V
430.00 V
1.450 mV
3.370 mV
13.570 mV
32.500 mV
66.100 mV
SC300 Option
6
Tolerance
(V p-p)
6-107