6-99. Equipment Required for Calibration and Verification

Table 6-41 lists the equipment, recommended models, and minimum specifications
required for each calibration and verification procedure.
Instrument
Digital
Multimeter
Adapter
Termination
BNC Cable
High-
Frequency
Digital Storage
Oscilloscope
Attenuator
Adapter
BNC Cable
AC
Measurement
Standard
Adapter
Termination
BNC Cable
Digital
Multimeter
Adapter
Termination
BNC Cable
6-74
Table 6-41. SC300 Calibration and Verification Equipment
Model
Wave Generator, Edge Amplitude Calibration, AC Voltage Verification
HP 3458A
Pomona #1269
(supplied with SC300)
Edge Rise Time and Aberrations Verification
Tektronix 11801 with
Tektronix SD-22/26
sampling head, or
Tektronix TDS 820 with
8 GHz bandwidth
Weinschel 9-10 (SMA)
or Weinschel 18W-10
or equivalent
(supplied with SC300)
Leveled Sine Wave Amplitude Calibration and Verification
Fluke 5790A
Pomona #1269
(supplied with SC300)
DC and AC Voltage Calibration and Verification, DC Voltage Verification
HP 3458A
Pomona #1269
(supplied with SC300)
Minimum Use Specifications
1.8 mV to
Voltage
Edge
4.5 mV to 2.75 V p-p Uncertainty: 0.06%
BNC(f) to Double Banana Plug
Feedthrough 50
1% (used with Edge Amplitude
Calibration and AC Voltage Verification)
Frequency
2 GHz
Resolution
4.5 mV to 2.75 V
10 dB, 3.5 mm (m/f)
BNC(f) to 3.5 mm(m)
Range
5 mV p-p to 5.5 V p-p
Frequency
50 kHz
BNC(f) to Double Banana Plug
Feedthrough 50
1%
BNC(f) to Double Banana Plug
Feedthrough 50
1%
105 V p-p Uncertainty: 0.06%