6-1. Introduction

This chapter contains the following information and service procedures for the
SC600 Oscilloscope Calibration Option functions.
Specifications
Theory of Operation
Calibration Procedures
Verification Procedures
Hardware Adjustments made after Repair
The calibration and verification procedures provide traceable results for all of the SC600
functions as long as they are performed using the recommended equipment. All of the
required equipment, along with the minimum specifications, is provided in Table 6-15
under "Equipment Requirements for Calibration and Verification."
The calibration and verification procedures in this chapter are not the ones Fluke uses at
the factory. These procedures have been developed to provide you with the ability to
calibrate and verify the SC600 at your own site if necessary. You should review all of the
procedures in advance to make sure you have the resources to complete them. It is
strongly recommended that, if possible, you return your unit to Fluke for calibration and
verification.
Hardware adjustments that are made after repair, at the factory or designated Fluke
service centers, are provided in detail.

6-2. Maintenance

There are no maintenance techniques or diagnostic remote commands for the SC600 that
are available to users. If your SC600 is not installed or not receiving power, the following
error message appears on the display when you press
calibration menus.
If this message is displayed, and you have the SC600 installed in your Calibrator
Mainframe, you must return the Calibrator Mainframe to Fluke for repair. If you wish to
purchase the SC600, contact your Fluke sales representative.
SC600 Option
to access the oscilloscope
6

Introduction

om030i.eps