3-15. Calibration Remote Commands

Calibration of the 5520A using remote commands is simple. To access the standard
calibration steps, simply send the command:
CAL_START MAIN
To jump to specific calibration steps, this command above can be modified by specifying
an entry point. The allowable entry points are as shown in Table 3-18.
Table 3-18. Jumping to a Specific Calibration Step in Remote
Entry points for CAL_START MAIN
AC Volts
Thermocouple Measuring
DC Current
AC Current
AUX DC Volts
AUX AC Volts
Resistance
Capacitance
Entry points for CAL_START FACTORY
NORMAL Volts and AUX Volts Phase
Volts and Current Phase
5500A
CALIBRATOR
NORMAL
AUX
SCOPE
V, , ,RTD
A, -SENSE, AUX V
OUT
PREV
STBY
OPR
EARTH
SCOPE
BOOST
MENU
HI
1000V
20V
1V PK
RMS
RMS
MAX
MAX
MAX
dBm
sec
LO
150V
7
8
9
m
V
Hz
PK
MAX
TRIG
20V
RMS
n
W
¡F
4
5
6
MAX
k
A
¡C
20V
p
1
2
3
M
F
20A
SHELLS
PK
GUARD
NOT
MAX
GROUNDED
/ +
0
SHIFT
ENTER
20V PK MAX
TC
20V PK MAX
AUX
SCOPE
0.1 Ohm shunt placed as closely as possible
OUT
to the AUX terminals of the 5520A
20V
RMS
MAX
150V
PK
MAX
TRIG
If the Phase Meter LO terminals are not common
20A
use a short between NORMAL LO and AUX LO
on the 5520A
Figure 3-16. Volts and Current Phase Verification
Precision
Phase
Meter
EDIT
SETUP
RESET
FIELD
NEW
CE
REF
MEAS
TRIG
TC
OUT
POWER
MULT
DIV
x
I
O
AV
TEMPX
ICAL
AI
V2
AVS
R
C
PHASE
IPHASE
Calibration and Verification

Calibration Remote Commands

CH2
CH1
yg015f.eps
Modifier
Modifier
3
3-27